This study aims to offer a voltammetric method for determining the antipsychotic drug flupentixol from serum samples. According to pH and scan rate studies, the oxidation behavior of flupentixol at boron-doped diamond electrode was found as irreversible and diffusion-controlled. In order to determine flupentixol from bulk form and serum samples, differential pulse voltammetry was preferred as the working method because of the repeatability. Linear responses were obtained in the range of 6.0×10-7 – 8.0×10-6 M and 8.0×10-7 – 1.0×10-5 M for bulk form and serum samples in pH 2.0 Britton-Robinson buffer solution, with detection limit values of 1.09×10-7 M and 1.08×10-7 M, respectively. Required validation parameters were also studied and according to recovery from serum samples (99.91 %) and precision studies, it can be said that developed method was accurate and precise.
Primary Language | English |
---|---|
Subjects | Analytical Chemistry |
Journal Section | Articles |
Authors | |
Publication Date | August 31, 2022 |
Submission Date | September 21, 2021 |
Acceptance Date | April 15, 2022 |
Published in Issue | Year 2022 Volume: 9 Issue: 3 |