The purpose of this study is to calculate self-attenuation correction factors, linear (LAC) /mass attenuation coefficients (MAC), half value layers (HVL) and tenth value layers (TVL) of different brands of commercial salt samples using gamma-ray spectrometry equipped with high resolution germanium (HpGe) detector. The gamma-rays emissions of 22Na, 60Co, 133Ba and 137Cs point sources were counted with/without sample. The obtained gamma-ray spectra were analyzed using computer software. Self-attenuation correction factors and gamma-ray attenuation parameters of eleven different brands of commercial salt sample were calculated. The experimental MACs of salt samples were compared with those of NaCl compound utilizing WinXCom software.
Self-attenuation correction factor gamma-ray attenuation parameter gamma-ray spectrometry salt samples
I would like to thank Dr. Christian SEGEBADE for useful suggestions and comments.
Primary Language | English |
---|---|
Subjects | Metrology, Applied and Industrial Physics |
Journal Section | Research Articles |
Authors | |
Publication Date | April 1, 2020 |
Submission Date | May 27, 2019 |
Acceptance Date | March 10, 2020 |
Published in Issue | Year 2020 Volume: 24 Issue: 2 |
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.