Amaç: Primer stabilite (PS), implantın uzunluğuna ve çapına, implant yüzeyinin mikro morfolojisine, implant tasarımına, cerrahi prosedüre, kemik miktarına ve kalitesine bağlıdır. Bu çalışmanın amacı, kısa ve standart dental implantları Tip IV kemikte primer stabilitelerine göre Rezonans Frekans Analizi (RFA) test ekipmanı kullanarak karşılaştırmaktır.
Gereç ve Yöntemler: Yetmiş iki dental implant; Large grit kumlanmış ve asitlenmiş (sandblasting with large grit particles and acid etching, SLA) (Dentium®, SimpleLine II, Seul, Kore), SLActive yüzey- (Straumann Roxolid®, Basel, İsviçre), Resorbable blast media (RBM) yüzey- (Implance®, Trabzon) (TÜRKİYE) kısa (4 ve 6 mm uzunluğunda) ve standart dental implantlar (10 mm uzunluğunda), insan tip IV kemiğine benzer kemik kalitesine sahip sığır türü kaburgaya yerleştirildi. Üç grubun parametreleri Mann-Whitney U test ile karşılaştırıldı.
Bulgular: Dört ve 6 mm uzunluktaki implantların İmplant Stabilite Katsayısı (Implant stability quotient, ISQ) ölçümleri benzerdi, ancak 10 mm uzunluktaki implantlarda gruplar arasında istatistiksel olarak anlamlı fark gözlendi. On mm uzunluğundaki implant grubunda SLActive yüzey en yüksek ISQ değerini gösterdi (p <0.005). Primer stabilite grup içinde karşılaştırıldığında, SLA yüzey ve RBM yüzey implantlarda önemli bir fark gözlenmezken 10 mm uzunluktaki SLActive yüzey implantı 4 ve 6 mm uzunluğundan daha yüksek ISQ değerine sahip bulundu (p < 0.005).
Sonuç: Kısa implantların Primer Stabilitesinde implantların uzunluğu ve çapı, yüzey tasarımına ve kısa implantlardaki yüzey özelliklerine kıyasla daha etkili olabilir ve SLActive yüzey implantları, standart implantlarda daha iyi primer stabilite gösterebilir.
Background:Primary stability (PS) is dependent on the length and diameter of the implant, micro-morphology of the implant surface, implant design, surgical procedure, bone quantity and quality.The aim of this study was to compare short and standart dental implants according to primary stability in bone type IV using Resonance frequency analysis (RFA) test equipment.
Methods: Seventy two dental implants; sandblasting with large grit particles and acid etching (SLA) surface (Dentium®,SimpleLine II, Seoul, Korea), SLActive surface- (Straumann Roxolid®, Basel, Switzerland), Resorbable blast media (RBM) surface- (Implance®,Trabzon, TURKEY) with short (4-and 6-mm length) and standart dental implants (10-mm length) were placed in bovine cow ribs with bone quality similar to a type IV human bone.RFA was performed to evaluate the primary stability. The parameters of three groups were compared with the Mann–Whitney U test.
Results:Implant stability quotient (ISQ)measurements of 4-mm and 6-mm length implants were similar however, 10-mm-length implants showed statistically difference inter-groups. SLActive surface in 10-mm-length implant group showed highest ISQ value (p < 0.005). When primary stability was compared in intra groups, SLA surface and RBM surface implants showed no significant difference, but 10-mm-length of SLActive surface implant showed higher ISQ value than 4-and 6-mm length (p< 0.005).
Conclusion: The length and diameter of the implants might be more effective in primary stability compared to surface design and surface properties in short implants and SLActive surface implants may show better primary stabiliy in standard implants.
Primary Language | English |
---|---|
Subjects | Dentistry |
Journal Section | Research |
Authors | |
Publication Date | August 1, 2020 |
Submission Date | March 11, 2019 |
Published in Issue | Year 2020 Volume: 7 Issue: 2 |
Selcuk Dental Journal is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License (CC BY NC).